Unit RP 1090 RP 2090 RP 10100 RP 290 E Pump level
at -60 °C W 370 340 320 390 4
at -70 °C W 240 180 250 210 4
at -80 °C W 110 70 170 90 4
at -90 °C W 20 10 60 10* 4
at -100 °C W --- --- 10 --- 4
Cooling water con-
nection sockets
Inches
Thread G3/8"
external &
G1/4" internal
Thread G3/8"
external &
G1/4" internal
Thread G3/8"
external &
G1/4" internal
Thread G3/8"
external &
G1/4" internal
---
Flow of cooling water
with temperature of
15 °C
l/min
1 bar ð 2.1 1 bar ð 2.1 1 bar ð 2.1 1 bar ð 2.1
---
* measured with pump level 2
11.3 Refrigerant and filling weight
The device contains fluorinated greenhouse gases.
Table 41: Single-stage refrigerating machine
Unit RP 2040* RP 3035* RP 2045*
Refrigerant --- R-404A R-404A R-404A
Maximum filling weight kg 0.25 0.25 0.3
GWP
(100a)
*
--- 3922 3922 3922
CO
2
equivalent
t 1.0 1.0 1.2
Unit RP 240 E RP 245 E RP 250 E
Refrigerant --- R-449A R-449A R-452A
Maximum filling weight kg 0.3 0.3 0.3
GWP
(100a)
*
--- 1397 1397 2140
CO
2
equivalent
t 0.4 0.4 0.6
* Only available outside the EU
Devices containing semi-halogenated
refrigerant
V6PRO bath thermostats and circulation thermostats136 / 156