CAUTION!
Risk of heat transfer liquid escaping from consuming unit in an
elevated location
Electric shock
If the external consuming unit is positioned above the
device, heat transfer liquid may escape from the device
when the pump stops. Therefore use the return protection
available as an accessory in the external hydraulic circuit.
CAUTION!
Risk of heat transfer liquid escaping from consuming units in a
higher or lower location
Electric shock
If the external consuming unit is positioned above or below
the device, heat transfer liquid may escape from the device
or consuming unit when the pump stops. Therefore use
the constant level device available as an accessory in the
external hydraulic circuit.
CAUTION!
Risk of external consuming unit bursting
Scalding, cold burns
If the external consuming unit is located in a lower position
and is sensitive to pressure, also take into account the
additional pressure resulting from the dierence in height
between the consuming unit and the device.
Please also note the following:
Always use the largest possible diameters and shortest possible hose
lengths in the external circuit.
If the hose diameter is too narrow, the insucient flow rate will cause
a drop in temperature between the device and the external consuming
unit. In this case, increase the bath temperature or pump level accord-
ingly.
Secure the hoses using hose clips.
V6 PRO bath thermostats and circulation thermostats 55 / 156

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lauda P 10 and is the answer not in the manual?

Lauda P 10 Specifications

General IconGeneral
BrandLauda
ModelP 10
CategoryThermostat
LanguageEnglish

Related product manuals