EasyManua.ls Logo

Launch X-431 PAD III - Trigger setting

Launch X-431 PAD III
161 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LAUNCH X-431 PAD III User Manual
100
reference lead to the ground connector. When using the probe hook-tip,
insert the tip onto the probe compensator firmly to ensure a proper
connection.
Fig 12-2
5. Tap the button located on the bottom of the screen, a square wave
(approximately 1kHzm 2V peak-to-peak) will be displayed within several
seconds.
*Note: The above steps also can be applied to check whether the signal
input/output of other Channels are normal or not.
Check the shape of the displayed waveform to determine whether the probe
is correctly compensated.
Correctly
Compensated
Over
Compensated
*Note: If necessary, use a
non-metallic tool to adjust the
trimmer capacitor of the probe for
the fattest square wave being
displayed on the Scopebox.
Under
Compensated
*Warning: To avoid electric shock while using the probe, make sure the insulated cable
is perfect, and do not touch the metallic portions of the probe head while it is
connected with a high-voltage source.

Table of Contents

Related product manuals