Leica DVM6 User Manual Troubleshooting 78
Troubleshooting (Continued)
Warnings
•
If the Leica DVM6 with motorized XY
stage is re-initialized, the stage moves
to the end position. Ensure that there are no
obstacles in the stage's range of movement.
•
The XY stage moves up when a
Leica DVM6 with a motorized focus
drive is restarted. Remove samples and other
objects from the XY stage to avoid damaging
the objective in use.
Error Correction
Uneven image brightness when
using the coaxial illumination
Set the relief contrast to the OPEN position. Otherwise there
may be a perceptible change in brightness within the image,
particularly at low magnification levels.
Set the contrast dial for the λ/4 plate to the Normal position.