-52-
Trigger Action Setting
Touch the setting display of trigger action and touch the trigger action
to be set.
■
Buzzer
Sounds the buzzer for about 0.3
second.
■
Stop measurement
It automatically stops measurement after the specified length of
measurement.
Quick
:
Stop immediately
Before
:
About 64k data
Center
:
Data of about the half
amount of capture
memory.
After
:
Data of a little bit less
than the amount of
capture memory
■
Save data
It saves up to 9999 numbers
of measured data of before the
trigger match (offset -) and after
the trigger match (oset +) into a
USB memory as “TRGSAVEnn.
DT” le.
(* 1)
While saving the next saving process will be ignored. If you stops the
measurement while processing the le will not be saved.
* 1 : The “nn” of le name will be allocated from 00 to 99 in the order of the saving and
if it passes 99 while measurement it returns to 00 and overwrite the old ones (which
have same “nn”). If you restart the measurement by [Run] the nn starts from 00 again
and overwrites the les which has the same “nn” number.