D9U001MCX-0101_03
53
Status (Pop-up window)
Meaning How to change the status
Positioning is disabled during activated
Automatic Lateral Therapy.
Deactivate the running Automatic
Lateral Therapy if bed positioning
is needed.
Lowering of the mattress support platform
is stopped because of a detected ob-
stacle on the undercarriage cover.
Remove the obstacle from the
undercarriage cover to continue
in lowering the mattress support
platform.
Automatic Lateral Therapy cannot be
activated because of overloaded bed.
Remove load to enable activation
of the Automatic Lateral Therapy.
Automatic Lateral Therapy cannot be
activated because of the Backrest in
more than 30°.
Lower the Backrest Angle to
enable activation of the Automatic
Lateral Therapy.
Automatic Lateral Therapy cannot be
activated because of too high Trendelen-
burg Angle or Antitrendelenburg Angle.
Lower the Trendelenburg Angle or
Antitrendelenburg Angle to enable
activation of the Automatic Lateral
Therapy.
Automatic Lateral Therapy cannot be
activated because of too low position of
the mattress support platform.
Adjust a higher position of the
mattress support platform to
enable activation of the Automatic
Lateral Therapy.
ALT unavailable - Bed overloaded.
Remove load to use ALT.
Positioning disabled during ALT.
Obstacle detected.
Remove the obstacle to lower the bed
platform.
ALT unavailable - Backrest >30°.
Lower the angle of backrest.
ALT unavailable - TR/ATR angle is
too high for seleceted ALT angle.
ALT unavailable - Bed platform too
low for selected ALT angle.