System Fault
Registered
OptiCare fault nding using displayed fault codes
Fault
Code 02
Fault
Code 04
Fault
Code 12
Fault code
14/15/16/
17
Analyse the
fault code
Power o󰀨 &
on the SCU &
when running
select CPR
Is the
SCU Mains
green light
on?
Is
mattress
LCD screen (or
display)
visible?
Are any
of the pipes
restricted?
Is the
CLP air pump
running?
Check
the serial
data cable for
disconnecti-
on or
damage
Remove the SCU for
bench testing & fault
nding
Check fault code displayed on iBoard Standard
or on Multiboard mattress screen
YES
NO
YES
Fault
Code 03
Fault
code 91
Fault
code
13/105
Fault
Code 01
Fault
code 93
Fault
Code 09
Are
the Red & Black
air connections
OK?
Are the
mattress internal
connections,
including manual
CPR, OK?
Are
mattress cells
undamaged?
CLP pressure
raised / lowered
& CLP icon
ashes
Replace / re-t
or repair
Can
you hear the
dump valve
operate?
Replace
/ Repair
Are any
of the pipes
restricted?
Are
any of the
pipes
kinked or
leaking?
Check the
auxiliary power
supply
Remove the SCU
for bench testing
& fault nding
Is the
bed auxiliary
power on?
Did the
solenoid valve
operate?
Did the
Fowler Boost
valve
operate?
Is the
mattress in
Fowler
mode?
Fault Log Self
Cleared
Power on
hours cleared
NO
NO
YES
YES
YES
YES
NO
NO
NO
YES
NO
NO
NO
YES
YES
NO
NO
YES
YES
NO
YES
NO
NO