3030226 Rev. D 5-20 Routine Maintenance
Materials Required:
• MINNCARE HD Disinfectant (select one)
- 4 x 1 Qts P/N 3029765
- 6 x90 mL P/N 3031105
• MINNCARE HD Residual Test Strips P/N 3029795
• Chemical Bottle (Provided with the RO) P/N 3026770
• “DO NOT USE” warning labels (not provided by Mar Cor Purification)
• Cleaning/Disinfect log sheets
• Current log sheets for water quality comparison
• PPE
NOTE:
Read and understand instructions before beginning procedure.
CAUTION:
MINNCARE HD is corrosive to most metals other than stainless steel. The
Millenium HX uses only compatible materials. MINNCARE HD contains oxidizing
agents that are potentially damaging to thin film (polyamide) membranes. It is
imperative that all the precautions and instructions are followed correctly. Failure
to comply with these instructions, which incorporate instructions from the
membrane manufacturer, will jeopardize the system’s performance and warranty.
1. Connect the water inlet line to the water supply. Connect the waste line to a
functional drain. Ensure that the product line is connected between the product
and return ports on the RO. Turn the RO and water supply on. Disconnect the
dialysis machine from the product line if connected.
2. Prepare the sanitizer by adding RO water and the amount of sanitizer as
specified on the labeling of the chemical bottle. Do not connect the bottle at
this time.
CAUTION:
Water to be used must be chlorine free or damage to the RO membrane may
result.
WARNING:
Label machine with appropriate warning signs such as "DO NOT USE CONTAINS
SANITIZER" to prevent use of the RO until it is properly rinsed.