1.6 Environmental Considerations
Prior to the installation of the Millenium HX RO, it will be necessary to provide utilities
and create an environment suitable for the trouble free operation of the RO and its
accessories.
WARNING:
Ordinary pipe dope must be avoided since it may leach objectionable and
potentially dangerous impurities into the water.
CAUTION:
It is necessary to keep the RO from freezing to prevent serious damage to the RO.
CAUTION:
The use of mobile telephones or communication equipment in the vicinity of the
Millenium HX could adversely influence the performance of the machine. See
specification in Section 4.8.
CAUTION:
Millenium HX needs special precautions regarding EMC and needs to be installed
and put into service according to the EMC information provided in the Operators
and Service Manuals.
1.6.1 Power
The RO operates on 115/230 VAC single-phase power. Histories of power failure,
power surges, and low line voltages should be noted and reported to the manufacturer
or their agent as they may create adverse conditions for the operation of equipment.
1.6.2 Water
Reverse osmosis systems require a reliable water source (rapid pressure fluctuation of
greater than 10 PSI should be avoided to prevent damage). The RO needs a minimum of
20 PSI pressure at a minimum flow of 1.5 gallons per minute. It is important that water
is flowing at the design flow rate when testing water pressure rather than reading static
pressure. Inlet water must contain less than 0.1 ppm free chlorine.
1.6.3 Drain
A drain outlet is required for the RO. The drain must have a minimum capacity of 2
gallons per RO per minute of continuous flow for as long as and as often as the RO is
operating.