3030226 Rev. D 3-8 System Operation
3.3.1 Heat Disinfection Procedure (Single Heat)
Materials Required:
• Cleaning/disinfection log
NOTE:
Read and understand instructions before beginning procedure.
1. Connect the water inlet line to the water supply. Connect the waste line to a
functional drain. Ensure that the product line is connected between the product
and return ports on the RO. Turn the RO and water supply on.
CAUTION:
Water to be used must be chlorine free or damage to the RO membrane may
result.
2. Press and hold the blue banner on the top of the Main Screen that says PRESS
AND HOLD FOR CLEANING MODES.
3. Press and hold the HEAT PROCESS SELECT button.