Millenium HX RO System
3030226 Rev. D 7-18 Drawings
7.4 Maintenance Schedule
As
Required
Per
Treatment
Day
Weekly
Monthly
Quarterly
Annually
Record hardness
(if applicable)
X
Verify salt level
(if applicable)
X
Record
chlorine/chloramine
levels
(per treatment)
X
Record %
rejection/water
quality
X
Record pre-filter
pressures
X
Disinfect the system
(minimum)
X
Bacteria/Endotoxin
test
X
Visually inspect
pre-filter
X
Clean RO membrane
X
Sanitize RO with
MINNCARE HD
X
Verify TDS/µS and %
Rejection
X
inspect the QD
fittings and lines
X
AAMI/CSA water
quality testing
X
RO Feed Water Test
X
Inspect tubing and
fittings for leaks or
wear
X
Inspect all pre RO
devices for correct
operation
X