Millenium HX RO System
3030226 Rev. D 3-11 System Operation
10. After the POST HEAT RINSE is complete, the next screen shows that the heat
process completed successfully. Press EXIT to move to the next screen.
11. Press EXIT to move to the Main Screen from the Chemical or Heating Operations
selection screen.
NOTE:
To further verify that the heat process was completed, press the LOGS button on
the Main Screen to confirm completion.
3.4 System Storage
This storage procedure has been established to store Millenium HX RO’s. It is not
recommended that a unit be dormant longer than 72 hours without disinfecting. An RO
that has been left unused for more than 72 hours risks the possibility of colonizing
bacteria and exceeding acceptable limits. A longer dormant period without disinfection
is the responsibility of the operating facility. RO’s to be stored should first be
disinfected with one of the procedures in this manual. The RO should be stored in a
location that would avoid freezing and excessive heat.
CAUTION:
MINNCARE HD is not acceptable for RO storage.