3030226 Rev. D 6-7 Troubleshooting
6.9 Service Required Component Failure Detected
The OPLC monitors various signals during operation to detect a failure of components.
This is active while the RO is running. Below are the error codes. Have this code ready
when you contact Technical Support.
RO Feed quality probe error, out of range, quality PCB circuit error, Fuse F5
Open
Product quality probe error, out of range, quality PCB circuit error, Fuse F5
Open
Product temperature probe error, out of range, quality PCB circuit error, Fuse
F5 Open
High product pressure during chemical (SV5 or SV3 may be faulty)
Major drop in water temperature during heat hold phase
Solenoid valve 4 error, tank filling, tank draining or flow problem
Pump fault during dialysis, chemical or cooling mode
Pump high pressure condition
Low product production in any disinfection/cleaning
RO Feed pressure transducer PS1 fault/error, Fuse F5 Open
Pump fault during the heat process
Pump pressure transducer PS2 fault/error, Fuse F5 Open
Product pressure transducer PS3 fault/error, Fuse F5 Open
Tank filling error, level switch error, water flow path error
Tanks level sensor error, tank level error during heat purge, tank thermal
protection switch trigger, analog component error
Membrane door or membrane incorrectly secured/installed
Thermal safety switch circuit opened
Tank level sensor sequence error