Millenium HX RO System
3030226 Rev. D 3-6 System Operation
9. Record data on a daily log sheet.
WARNING:
The RO should not be placed in service until data on the log sheets meet the
minimum standards set at your facility.
10. The Millenium HX RO is now ready for connection to the dialysis machine (if not
already connected).
3.2 Shut-Down Procedure
1. From the RUN screen, press and hold the STOP pushbutton for 2 seconds. The
RO will coninue to operate for 60 seconds in a shutdown flush. When in shut
down flush, the RUN button is not visible until the process is complete. The RO
may be heat disinfected at this time.
NOTE:
If the RO is going to be off line for an extended period of time, Mar Cor
Purification recommends that the RO be setup for storage heat disinfection. Refer
to Section 3.4.1 Storage Heat Procedure. The operator can also perform a single
heat. Refer to Section 3.3.1 Heat Disinfection Procedure (Single Heat).
2. Turn off the inlet water supply.
3. Disconnect from the utilities and store the RO.