3030226 Rev. D 4-20 Screen Operation And Specifications
4.4 Alarm Screens
This screen is displayed indicating that a component within the
RO has failed along with a possible error code. Refer to Section
6.9 “Service Required Component Failure Detected” for the error
code list.
This device has a tilt sensor to prevent operation in a horizontal
orientation. If the device is placed on its side, operation of the
device ceases and it cannot be started until the RO is placed in an
upright position.
This device has a product water temperature sensor that is
integral to the conductivity probe. If the product water
temperature exceeds 105°F (41°C), the RO will shut down and
cannot be restarted in run until the product water sensor is stable
below 105°F (41°C).
This device has a pressure transducer that monitors the pump
outlet pressure. If pump pressure exceeds 240 PSI, or is 20 PSI
below the target pressure, the RO will shut down. The pressure
alarm settings are not user adjustable.
During the heat process, pump pressure is monitored and if the
pressure goes outside of a factory set range, the Pump Pressure
Alarm will sound. These settings are not user settable.