3030226 Rev. D 5-19 Routine Maintenance
5.4 Chemical Sanitization Procedure
NOTE:
Chemical disinfection is not required for Millenium HX units. Heat disinfection is
the preferred method to maintain bacterial control.
WARNING:
Always wear rubber gloves, apron and facemask. Use proper ventilation during
disinfection and rinse-out. DO NOT BREATHE FUMES OR ALLOW MINNCARE HD
TO COME IN CONTACT WITH SKIN OR EYES. Always follow recommendations
found in the Material Safety Data Sheet and any applicable OSHA standards for
the chemical being used.
WARNING:
Label machine with appropriate warning signs such as “DO NOT USE/CONTAINS
SANITIZER” to prevent use of the RO until it is properly rinsed.
CAUTION:
The different cleaners/sanitizers must not be mixed together.
NOTE:
MINNCARE HD STORAGE AND DISPOSAL:
Do not contaminate water or food by storage or disposal.
STORAGE:
Store in shipping carton. Do not expose to direct sunlight. Maintain temperature
below 75°F (24°C). Avoid contact with combustible materials. Avoid
contamination from any source, including metals, dust, etc. Such contamination
may cause rapid decomposition, generation of large quantities of oxygen gas and
high pressure. Store in original closed container. NEVER TAMPER WITH VENT.
DISPOSAL:
Waste resulting from the use of this product may be disposed of on-site by
dilution in a sanitary sewer or at an approved waste disposal facility.
CONTAINER DISPOSAL:
Triple rinse empty container with water, then offer for recycling or reconditioning
or puncture and dispose of in a sanitary landfill; incinerate; or if allowed by state
and local authorities by burning. If burned, stay clear of the smoke.