3030226 Rev. D 6-4 Troubleshooting
6.3 Product Water Quality Alarm
1. Product Quality Shutdown during normal run.
a. The Product water Quality set point is adjustable from 1 to 50 TDS (1-100
µS). Verify that your set point is correctly set.
6.4 This Unit Must Stand Upright Alarm
1. This Unit Must Stand Upright shutdown.
a. Verify the Unit is in the upright position.
b. With the unit disconnected from electricity, verify that the tilt sensor is
closed using an ohmmeter when in the upright position. If faulty, replace tilt
sensor.