3030226 Rev. D 5-29 Routine Maintenance
9. Plug the power cord into a 115 VAC (230 VAC) grounded outlet and reset the
GFCI if necessary (60 Hz RO only). Do not use an extension cord or an un-
grounded adapter.
10. Turn on the water supply to the RO.
11. Turn on the main power module power switch. The touch screen will illuminate.
Press START RINSE PROCESS to rinse RO. Allow the RO to rinse for 30 steps.
12. Press EXIT to enter the Chemical or Heating Operations selection screen after
the rinse is complete.
13. Complete a SINGLE HEAT disinfection. Refer to Chapter 3, Section 3.3.1 “Heat
Disinfection Procedure (Single Heat)”.