3030226 Rev. D 1-4 General Information
WARNING:
Only chemicals listed in the manual may be used with the Millenium HX RO. Other
chemicals may not be compatible with the materials of construction. When using
chemical cleaners or sanitizers, ensure the system is properly rinsed prior to use for
dialysis.
WARNING:
The selection of water treatment equipment for dialysis is the responsibility of the
user and the product water should be tested periodically in accordance with AAMI
and CSA standards.
WARNING:
This device will reach temperatures up to 180°F (82°C). Only appropriate heat
resistant materials can be used for fluid pathways that are disinfected with hot
water.
WARNING:
After installation and subsequent use, if any device in the water treatment system is
changed or replaced, the user should conduct appropriate tests to ensure that the
revised system meets the initial design criteria.
WARNING:
The product water quality is proportional to the RO feed water quality. The product
water may not meet quality limits if RO feed water quality deteriorates significantly.
The user is responsible for monitoring the RO feed water quality.
WARNING:
Carbon cannot be regenerated and must be replaced with new beds when
exhausted or contaminated.
WARNING:
A suitable carbon filter should always be present as pretreatment for the RO in
order to prevent chlorine/chloramine damage to the thin film RO membrane and
harm to the patient. Mar Cor Purification strongly recommends the use of two
carbon devices, used in a series configuration.
WARNING:
This RO can only operate in the vertical position. This machine has been equipped
with a sensor that will prevent operation of the unit while lying horizontal. Damage
to equipment may result if operated in the horizontal position.