Millenium HX RO System
3030226 Rev. D 1-6 General Information
WARNING:
Do not remove the back panel of the RO during operation. System is under
pressure during operation. Only remove panel when RO is turned OFF.
WARNING:
To avoid physical and/or equipment harm, move the RO by using both handles at all
times.
CAUTION:
Do not cover or enclose the RO during operation to ensure the unit properly
dissipates heat.
CAUTION:
The input water quality to the RO must meet the National Primary Drinking Water
Standards of the United States Environmental Protection Agency (EPA) or Canadian
Federal and Provincial guidelines governing drinking water in Canada. If the input
water does not meet these standards, the RO may not produce water that meets
AAMI and CSA standards.
NOTE:
Use proper wiring and connection methods to satisfy hospital electrical codes.
NOTE:
Use proper water and drain connections to prevent contamination of the RO and
danger to the patient. The Product/Return Ports are male quick disconnect to
prevent mix up of connection ports.
NOTE:
The Millenium HX RO’s hot water disinfection process is effective in reducing at
least 99.9% of an introduced bacteria load of Pseudomonas aeruginosa under
normal operation conditions.
NOTE:
Clean the HMI display screen as necessary with an alcohol based fluid and a soft
towel as required to maintain the screen’s cleanliness.
NOTE:
This device does not supply hot water to disinfect hemodialysis machines.