3030226 Rev. D 3-16 System Operation
12. Once the CANCEL or ABORT button is pressed, the RO will exit from the Storage
Heat and display one of the following screens depending on the product water
temp.
NOTE:
The RO may be powered off, unplugged and relocated to a location where a drain
and RO feed water are available. When power is returned the RO will continue to
operate as described in the steps below.
a. If the RO is above 120°F (49°C): the RO will enter a slow cool process. If
the drain and RO feed water lines are connected, a FAST COOL can be
activated by pressing the green button. When 120°F (49°C) is reached,
the RO will proceed to the POST HEAT RINSE below.
b. If the RO is at or below 120°F (49°C): the RO will enter the POST HEAT
RINSE screen.
13. Connect the RO to water and drain (if not already connected) and press START
to begin the post heat rinse.