Rad-97 Chapter 11: Specifications
www.masimo.com 158 Masimo
Guidance and Manufacturer's Declaration-Electromagnetic
Emissions
Guidance and Manufacturer's Declarations - Electromagnetic Emissions
The ME Equipment is intended for use in the electromagnetic environment specified below.
The customer or the user of the ME Equipment should assure that it is used in such an
environment.
Emission Test Compliance Electromagnetic Environment - Guidance
RF Emissions
CISPR 11
Group 1
ME Equipment uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
RF Emissions
CISPR 11
Class B
Suitable for use in all establishments, including domestic
environments and those directly connected to the public
low-voltage power supply network that supplies buildings
used for domestic purposes.
Harmonic
Emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/
Flicker emissions
IEC 61000-3-3
Complies
Guidance and Manufacturer's Declaration-Electromagnetic
Immunity
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The ME Equipment is intended for use in the electromagnetic environment specified below.
The customer or the user of the ME Equipment should assure that it is used in such an
environment.
Immunity Test IEC
60601
Test Level
Compliance
Level
Electromagnetic Environment -
Guidance
Electrostatic
discharge (ESD)
+6 kV
contact
+6 kV contact
+8 kV air
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative

Table of Contents

Other manuals for Masimo Rad-97

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Masimo Rad-97 and is the answer not in the manual?

Masimo Rad-97 Specifications

General IconGeneral
Display Size7 inches
SpO2 Accuracy±2% (70-100%)
Pulse Rate Accuracy±3 bpm
Perfusion Index Range0.02% to 20%
Pleth Variability Index Range0% to 100%
SpO2 Range0% to 100%
SpHb Range0 to 25 g/dL
SpCO Range0% to 40%
Battery TypeRechargeable lithium-ion
Operating Temperature5°C to 40°C (41°F to 104°F)
Storage Temperature-20°C to 60°C (-4°F to 140°F)
Humidity Range10% to 95% non-condensing
DisplayColor touchscreen
ConnectivityWi-Fi, Ethernet, Bluetooth
Power Supply100–240 VAC, 50–60 Hz
Battery LifeApproximately 4 hours
MeasurementsSpO2, Pulse Rate, Perfusion Index, Pleth Variability Index, SpHb, SpOC, SpMet, SpCO
Pulse Rate Range25 to 240 bpm
HbCO Range0-99%
MetHb Range0-99%
SpMet Range0-99%
Data StorageInternal memory
AlarmsAudible and visual alarms for all parameters