EasyManuals Logo

Masimo Rad-97 Operator's Manual

Masimo Rad-97
188 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #152 background imageLoading...
Page #152 background image
Rad-97 Chapter 11: Specifications
www.masimo.com 150 Masimo
Parameter Resolution
SpMet 0.1%
SpOC 1.0 ml/dL
RRp** 1 RPM
* Breaths per minute is denoted as rpm on device display.
** Parameter currently not available in the U.S.A. and territories relying on FDA market
clearance.
Noninvasive Blood Pressure (NIBP) Specifications
Measurement Range
Measurement Patient Population Display Range
Systolic Adult 40-260 mmHg
Pediatric 40-230 mmHg
Neonatal 40-130 mmHg
Diastolic Adult 20-200 mmHg
Pediatric 20-160 mmHg
Neonatal 20-100 mmHg
MAP Adult 26-220 mmHg
Pediatric 26-183 mmHg
Neonatal 26-110 mmHg
Pulse Rate (PR) All 30-220 bpm

Table of Contents

Other manuals for Masimo Rad-97

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Masimo Rad-97 and is the answer not in the manual?

Masimo Rad-97 Specifications

General IconGeneral
Display Size7 inches
SpO2 Accuracy±2% (70-100%)
Pulse Rate Accuracy±3 bpm
Perfusion Index Range0.02% to 20%
Pleth Variability Index Range0% to 100%
SpO2 Range0% to 100%
SpHb Range0 to 25 g/dL
SpCO Range0% to 40%
Battery TypeRechargeable lithium-ion
Operating Temperature5°C to 40°C (41°F to 104°F)
Storage Temperature-20°C to 60°C (-4°F to 140°F)
Humidity Range10% to 95% non-condensing
DisplayColor touchscreen
ConnectivityWi-Fi, Ethernet, Bluetooth
Power Supply100–240 VAC, 50–60 Hz
Battery LifeApproximately 4 hours
MeasurementsSpO2, Pulse Rate, Perfusion Index, Pleth Variability Index, SpHb, SpOC, SpMet, SpCO
Pulse Rate Range25 to 240 bpm
HbCO Range0-99%
MetHb Range0-99%
SpMet Range0-99%
Data StorageInternal memory
AlarmsAudible and visual alarms for all parameters