Rad-97 Chapter 2: Description
www.masimo.com 37 Masimo
Back View
1. Nurse Call Connector
Allows connection to a Nurse Call system.
Caution: To ensure patient electrical isolation, all
external device connections to the Analog
Output/Nurse Call connectors must be IEC 60950-1,
IEC 60601-1, or UL 1069 compliant.
See Nurse Call Connection on page 41.
2. Ethernet
Allows network connection to Rad-97 using an RJ-45
cable.
3. USB
Provides USB 2.0 connectivity.
4. Equipotential Ground Connector
Provides optional functional earthing for Rad-97 to
eliminate potential differences between the earth
connections for Rad-97 and another medical device.
The use of the Equipotential Ground Connector
should be in accordance with IEC 60601-1.
5. Power Entry Module
Provides connection to an AC power cord.
Note: Always connect the Rad-97 to the mains power
for continuous operation and/ or battery recharging.
Note: Use the power cord as the means to disconnect
the device from AC power. To disconnect the device
from AC power, first disconnect the power cord from
the power outlet, rather than from the device.

Table of Contents

Other manuals for Masimo Rad-97

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Masimo Rad-97 and is the answer not in the manual?

Masimo Rad-97 Specifications

General IconGeneral
Display Size7 inches
SpO2 Accuracy±2% (70-100%)
Pulse Rate Accuracy±3 bpm
Perfusion Index Range0.02% to 20%
Pleth Variability Index Range0% to 100%
SpO2 Range0% to 100%
SpHb Range0 to 25 g/dL
SpCO Range0% to 40%
Battery TypeRechargeable lithium-ion
Operating Temperature5°C to 40°C (41°F to 104°F)
Storage Temperature-20°C to 60°C (-4°F to 140°F)
Humidity Range10% to 95% non-condensing
DisplayColor touchscreen
ConnectivityWi-Fi, Ethernet, Bluetooth
Power Supply100–240 VAC, 50–60 Hz
Battery LifeApproximately 4 hours
MeasurementsSpO2, Pulse Rate, Perfusion Index, Pleth Variability Index, SpHb, SpOC, SpMet, SpCO
Pulse Rate Range25 to 240 bpm
HbCO Range0-99%
MetHb Range0-99%
SpMet Range0-99%
Data StorageInternal memory
AlarmsAudible and visual alarms for all parameters