14) Set the adjustment and calibration parameters:
Perforated lid type: Not selectable. Can only be set during adjustment.
Run time check
(1)
: ON / OFF
15) Press the Start button to initiate the procedure, or
Press the Stop button to abort the process.
16) Once the process is finished the Website will be updated with the result:
Note
The maximum allowed Deviation from target air flow is ±4.0 %.
(1)
If activated, this feature compares the air samplers internal CPU clock with the PC’s watch. An error will be generated and logged in the air samplers
error log if the deviation is out of specification.