Page 67 of 188
User Manual MAS-100 Regulus en.docx, Version: 4.0
The air sampler must be re-adjusted according to chapter 7.5, chapter 7.4, chapter 8.4
or chapter 8.5 if the calibration result is outside the specified range.
17) To create a calibration certificate and print the displayed data afterwards, the user
has two options:
– As found certificate (continue with step 18), or
– As found certificate (continue with step 26)
Note
Upon a failed calibration, only the As found certificate button will be se-
lectable.
18) Press the As found certificate button.
19) Press on the hyperlink Go to the as found certificate at the bottom of the result page.
20) A new tab will open in the brower.
Press Print this page to either print the certificate on a printer or virtual PDF printer.