User Manual MAS-100 Regulus en.docx, Version: 4.0
21) Continue with step 30.
22) Set the calibration validity parameters:
Calibration reminder: ON / OFF
Calibration validity: 112months
Block measurements if calibration is expired
(2)
: ON / OFF
23) Press the As left certificate button.
24) Press on the hyperlink Go to the as left certificate at the bottom of the result page.
25) A new tab will open in the brower.
Press Print this page to either print the certificate on a printer or virtual PDF printer.
(2)
If activated, this feature will make it impossible to run a sampling if the calibration of the air sampler has expired.