Medin NC3 - Connecting the patient tube circuit set - High Flow

Medin NC3
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
medin Medical Innovaons GmbH
Adam-Geisler-Str. 1
82140 Olching, Germany
+49 8142 448 460
info@medin-medical.com
www.medin-medical.com
IFU_medin-NC3 EN
Rev 05 Status as of 28.04.2021
26 of 68
Warning:
The use of the Medijet®, the suitable Nuow® nasal cannula, the humidier and the paent tube kits is described
in the user manuals for these devices, along with important addional informaon and warnings regarding their
use. Users must also know and comply with these other user manuals, even though they may be familiar with the
medin-NC3® user manual.
Only suitable tube systems with appropriate adapters should be used to connect the paent and the Medijet® to
the medin-NC3®/the corresponding Nuow® nasal cannula, since otherwise there will be an increased likelihood
of disconnecons.
The humidier should be installed lower than the paent.
Connecng the paent tube circuit set – High Flow:
When High Flow mode is used, a suitable Nuow® nasal cannula must be connected instead of a Medijet®, following
mode selecon. The system set-up is shown in Figure 4-5.
INT
EXT
medical innovaons
Humidifier
Nuflow
Figure 4-5: Set-up of a medin-NC3® system with Nuow® nasal cannula