External baery
temperature
Low 90 s The temperature of the external
baery is too high.
Plug the power
supply unit in and
remove the ext.
baery.
Error message text Priority ∆t
(1)
Cause and device response Correcon
Pure alarm sound,
medin-NC3® switched o
- < 1 s This alarm is triggered if the
medin-NC3® power supply fails
completely, for example because
the power supply unit was not
connected to the mains although
the baery was discharged.
The medin-NC3® was switched o.
Plug in the power
supply unit and
restart the device
or contact a
service technician.
Press alarm reset
to restart ow
- - No alarm – informaonal message
only:
Triggered by a prolonged high
feed pressure (e.g. if a ow
tube is blocked), by several high
CPAP pressures successively
or by a combinaon of high
feed pressures and high CPAP
pressures, the ow supply is
interrupted for the me being
and is only reacvated aer the
Alarm Reset buon is pressed.
Reacvaon of the ow takes
approximately 10 seconds starng
from this moment.
Please check
paent tube for
blockage.
Push acvated info - Not actually an alarm, just
nocaon that a push
was triggered. The user can
decide whether to receive this
informaon visually or aurally.
-
Apnea – push acvated info - Not actually an alarm, just a
nocaon that a push was
triggered due to apnea. The user
can decide whether to receive this
informaon visually or aurally.
-
medin Medical Innovaons GmbH
Adam-Geisler-Str. 1
82140 Olching, Germany
+49 8142 448 460
info@medin-medical.com
www.medin-medical.com
IFU_medin-NC3 EN
Rev 05 Status as of 28.04.2021
53 of 68
Table 6: Low-priority alarms
(1)
Delay ∆t: This corresponds to the maximum me which passes from the me the alarm condion occurs
unl the alarm is triggered.