Backup push info -
Not actually an alarm, just a
nocaon that a backup push
was triggered due to apnea. Users
can decide whether or not to
receive this informaon aurally.
However, this informaon is
always indicated visually.
-
Type of alarm Volume in (dB)
In the case of
A-weighted background
noise volume (dB)
High priority (red) 66 ± 3 53 ± 2
Medium priority (blinking yellow) 66 ± 3 53 ± 2
Low priority (steady yellow) 66 ± 3 53 ± 2
Informaon signal (push) 66 ± 3 53 ± 2
medin Medical Innovaons GmbH
Adam-Geisler-Str. 1
82140 Olching, Germany
+49 8142 448 460
info@medin-medical.com
www.medin-medical.com
IFU_medin-NC3 EN
Rev 05 Status as of 28.04.2021
54 of 68
Table 7: Other alarms and messages
(1)
Delay ∆t: This corresponds to the maximum me which passes from the me the alarm condion occurs
unl the alarm is triggered.
The alarm volume is:
The funconality of the alarm signal generator is checked at each system start-up. The user must conrm that
the alarm sound can be heard.
Alterable alarms:
The user can set the CPAP pressure low, CPAP pressure high (yellow) alarms and the permied deviaon interval
for the Oxygen concentraon high / low alarms. This can be done using the seng in the Alarm Sengs menu
(see chapter 5.13). Aer the numerical values are adjusted, the modied sengs become acve and are saved
when this change is conrmed by pressing the dial. No other alarms can be modied.
Alarm presengs at device start-up:
Upon start-up, the CPAP pressure high (yellow) alarm is set to 12 mbar and the CPAP pressure low alarm is set to
3 mbar. The alarm interval for monitoring the oxygen levels set is always set to ± 5% at start-up.
If the medin-NC3® is restarted, the alarms are reset to these levels. This cannot be changed by the user.
The alarms CPAP pressure low (yellow) and CPAP pressure high (yellow) analyze whether the average value of the
CPAP readings is outside of the alarm thresholds (can be set under Alarm Sengs – see chapter 5.13). Thus the
alarm CPAP pressure high is triggered if the CPAP average value was above the alarm threshold for more than
one second. The alarm CPAP pressure low is triggered if the CPAP average value was below the alarm threshold
for more than 15 seconds.
Since the CPAP pressure low and CPAP pressure high alarms refer to the average CPAP pressure, the start and end
of the alarm is thus further delayed (see table at the start of the chapter). This likewise causes the CPAP pressure
low alarm to not be reset unl the CPAP average value is again above the alarm threshold and not directly aer
adjusng the alarm threshold or correcng the problem.