61
OPERATION
5. Read the water activity using the AQUALAB 3 (Section2.5) and make a note of the water
activity value.
This will be used in step13.
NOTE: The moisture content value has not yet been properly offset and should be ignored.
6. Use the reference method to determine the average moisture content of the other
twosubsamples.
This average is the reference moisture content value. This will be used in step14.
7. Open skalacontrol.com.
NOTE: Linear offsets must be performed on a product-by-product basis in the SKALA Control desktop
application. This functionality does not appear in the SKALA Mobile app.
Figure81 SKALA Main Menu
8. Select Products from the SKALA Main menu.
9. Select the Product group containing the product to be offset (Figure82).
Figure82 Product group
10. Click the edit icon next to the product to be offset (Figure82).
11. Make sure the model has been selected or set as instructed in (Section2.5.4).