EasyManuals Logo

METER AQUALAB 3 User Manual

Default Icon
90 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #65 background imageLoading...
Page #65 background image
61
OPERATION
5. Read the water activity using the AQUALAB 3 (Section2.5) and make a note of the water
activity value.
This will be used in step13.
NOTE: The moisture content value has not yet been properly offset and should be ignored.
6. Use the reference method to determine the average moisture content of the other
twosubsamples.
This average is the reference moisture content value. This will be used in step14.
7. Open skalacontrol.com.
NOTE: Linear offsets must be performed on a product-by-product basis in the SKALA Control desktop
application. This functionality does not appear in the SKALA Mobile app.
Figure81 SKALA Main Menu
8. Select Products from the SKALA Main menu.
9. Select the Product group containing the product to be offset (Figure82).
Figure82 Product group
10. Click the edit icon next to the product to be offset (Figure82).
11. Make sure the model has been selected or set as instructed in (Section2.5.4).

Other manuals for METER AQUALAB 3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the METER AQUALAB 3 and is the answer not in the manual?

METER AQUALAB 3 Specifications

General IconGeneral
BrandMETER
ModelAQUALAB 3
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals