EasyManuals Logo

Metrohm 883 Basic IC plus User Manual

Metrohm 883 Basic IC plus
107 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #17 background imageLoading...
Page #17 background image
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
883 Basic IC plus
■■■■■■■■
9
3
Installing the eluent path
Assemble the eluent aspiration tubing (6.1834.080) (3-1) and
connect it with the eluent bottle (see chapter 3.8, page 23).
Remove the stopper of connector 5 of the injection valve and
connect the column inlet capillary (6.1831.100) (3-2) using a PEEK
pressure screw (6.2744.014).
Connect the column inlet capillary (6.1831.100) (3-2) and the
capillary of the suppressor (3-4) labeled in to each other using a
coupling (6.2744.040) and 2 short pressure screws (6.2744.070).
The coupling is installed in the place of the separation column,
which may not be installed before the initial start-up.
Connect the capillary of the suppressor (3-5) labeled out and the
detector inlet capillary (3-6) to each other using a coupling
(6.2744.040) (3-18) and two short pressure screws (6.2744.070)
(3-15) (see "Connecting the detector input capillary to the sup-
pressor", page 44).
4
Installing the sample path
If you do not use a Sample Processor:
Guide the end of the sample aspiration capillary (3-19) out of the
instrument to the sample vessel and fasten it there.
Connect the end of the sample outlet capillary (3-20) from the
inside to the Luer connector of the door (6-1) using the PEEK
pressure screw (6.2744.070) (6-3). The sample can then be aspi-
rated from outside with a syringe.
If you use a Sample Processor:
Guide both capillary ends out of the instrument.
Connect the sample aspiration capillary with the Sample Processor
(see the manual for the Sample Processor).
Guide the sample outlet capillary into a waste container and fas-
ten it there.
5
Installing the peristaltic pump
(see chapter 3.14, page 38)
Connect the capillary (6.1803.020) used as aspiration capillary for
the regeneration solution (3-8) to the aspiration end of the pump
tubing (6.1826.320) (3-9) using a tubing olive (6.2744.034)
(3-16) and a short pressure screw (6.2744.070) (3-15). Shorten it
to the required length.
Place the pump tubing in a tubing cartridge.
Insert the tubing cartridge into the peristaltic pump.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 883 Basic IC plus and is the answer not in the manual?

Metrohm 883 Basic IC plus Specifications

General IconGeneral
BrandMetrohm
Model883 Basic IC plus
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals