4 Operation and maintenance
883 Basic IC plus
■■■■■■■■
63
3
Reinserting the outlet valve into the pump head
■ Insert the outlet valve into the outlet valve holder (the seal must
be visible) (1).
■ Screw the outlet valve holder up into the pump head (2) and
tighten it firmly by hand and then retighten it one additional ¼
turn using the adjustable wrench (3).
■ Tighten the connection capillary to the auxiliary piston back onto