4.13 Metrohm CO
2
Suppressor (MCS)
■■■■■■■■■■■■■■■■■■■■■■
46
■■■■■■■■
930 Compact IC Flex Oven/SeS/PP (2.930.2500)
NOTE
The new CO
2
adsorption cartridge CW (6.2837.100) works without
attached H
2
O adsorption cartridge.
Installing the CO
2
adsorption cartridge CW
Accessories ■ Prepared CO
2
adsorption cartridge CW (6.2837.100)
CAUTION
The following preparatory steps absolutely must be carried out for CO
2
suppression to operate correctly.
Install the CO
2
adsorber cartridge CW as follows:
1
Connecting the CO
2
adsorber cartridge CW
Attach the capillary connected to the Metrohm CO
2
Suppressor's
(MCS) Air in connector to the tip of the CO
2
adsorber cartridge CW.
2
Placing the adsorber cartridge in the instrument
■ Place the CO
2
adsorber cartridge CW into the instrument's detec-
tor chamber.