4 Installation
930 Compact IC Flex Oven/SeS/PP (2.930.2500)
■■■■■■■■
51
3
Starting the equilibration
■ Start the equilibration in MagIC Net: Workplace ▶ Run ▶ Equi-
libration ▶ Start HW.
4
Deaerating the high-pressure pump
■ Push the end of the purge needle (6.2816.040) over the end of
the purge capillary on the purge valve.
■ Insert the syringe (6.2816.020) in the Luer connector of the purge
needle.
■ Open the purge valve using the rotary knob (approx. ½ turn).
■ Switch on the high-pressure pump in MagIC Net.
■ Use the syringe to aspirate eluent until there are no more air bub-
bles in the eluent aspiration tubing.
■ Switch off the high-pressure pump in MagIC Net.
■ Seal the purge valve using the rotary knob.
■ Remove the syringe from the purge needle.
■ Pull the purge needle out of the purge capillary.
5
Adjusting the contact pressure of the peristaltic pump
■ Adjust the contact pressure of the peristaltic pump (see "Setting
the contact pressure correctly", page 41).