4 Installation
930 Compact IC Flex Oven/SeS/PP (2.930.2500)
■■■■■■■■
51
3
Starting the equilibration
Start the equilibration in MagIC Net: Workplace Run Equi-
libration Start HW.
4
Deaerating the high-pressure pump
Push the end of the purge needle (6.2816.040) over the end of
the purge capillary on the purge valve.
Insert the syringe (6.2816.020) in the Luer connector of the purge
needle.
Open the purge valve using the rotary knob (approx. ½ turn).
Switch on the high-pressure pump in MagIC Net.
Use the syringe to aspirate eluent until there are no more air bub-
bles in the eluent aspiration tubing.
Switch off the high-pressure pump in MagIC Net.
Seal the purge valve using the rotary knob.
Remove the syringe from the purge needle.
Pull the purge needle out of the purge capillary.
5
Adjusting the contact pressure of the peristaltic pump
Adjust the contact pressure of the peristaltic pump (see "Setting
the contact pressure correctly", page 41).

Table of Contents

Other manuals for Metrohm 930 Compact IC Flex

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 930 Compact IC Flex and is the answer not in the manual?

Metrohm 930 Compact IC Flex Specifications

General IconGeneral
BrandMetrohm
Model930 Compact IC Flex
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals