6.7 Servicing the high-pressure pump
■■■■■■■■■■■■■■■■■■■■■■
72
■■■■■■■■
930 Compact IC Flex Oven/SeS/PP (2.930.2500)
1
Release the coupling from the pressure screw and seal it with a stop-
per.
2
Unscrew and remove the pressure screw on the pump head's outlet
(17-8).
3
Loosen and remove the four fastening screws (17-3) using the hex
key.
4
Remove the pump head (17-4).
Servicing the piston
Carry out the following work on both pistons in turn.
Servicing a piston consists of the following tasks:
1. Replace the piston seal.
2. Clean or replace the zirconium oxide piston.
3. Reinstall the piston.