AccuPyc II 1340 System Commands
Aug 2013 3-17
Calibration Data
The sample cell volume and the expansion cell volume are used for calculating the sample volume.
The cell volume and expansion volume are updated automatically when the pycnometer is calibrated.
These prompts enable you to view current calibration data.
Press
CHOICE until Calibration Data is displayed, then
press
ENTER.
Displays when you select Yes at the Request insert prompt
in Report Options.
Choices: None, 0.1 cm
3
(for 1 cm
3
pycnometer)
None, 3.5 cm
3
, 1.0 cm
3
(for 10 cm
3
pycnometer)
None, 35 cm
3
, 10 cm
3
(for 100 cm
3
pycnometer)
None, 650 cm
3
, 1300 cm
3
(for 2000 cm
3
pycnometer)
Refer to Calibrating Inserts, page J-7 for additional
information.
Press
CHOICE until the appropriate insert is displayed, then
press
ENTER.
Displays the cell volume determined during calibration.
Range: 0.01 to 999.0000 cm
3
Press ENTER.
Displays the expansion volume determined during
calibration.
Range: 0.5 to 999.0000 cm
3
Press ENTER.
Unit[n] SN1234 10 cm3
Set-up Type?
Calibration Data
Unit[n] SN1234 10 cm3
Calibration Data
Chamber Insert?
None
Unit[n] SN1234 10 cm3
Calibration Data
Cell Volume:
(value from calibration)
Unit[n] SN1234 10 cm3
Calibration Data
Expansion Volume:
(value from calibration)