System Commands AccuPyc II 1340
3-18 Aug 2013
Communications
Communications enables you to specify communications criteria, such as baud rate, parity, etc.
Press
CHOICE until Communications is displayed, then
press
ENTER.
This prompt enables you to choose the manner in which
your IP address is assigned. With an IP address, you can
access a web browser to view analysis results. Typically,
after an analysis (or calibration) is started, you cannot
review results from the previous operation. However,
results from five analyses are stored in the controller. A web
browser enables you to view them. (Refer to Using a Web
Browser, page 4-9 for step-by-step instructions.) An IP
address must exist for this operation. You can have an IP
address detected and assigned automatically, or you can
specify one.
Choices: DHCP, Specify
Be sure your ethernet cable is plugged into the Ethernet port
on the rear panel of the analyzer and to the network before
proceeding with this option.
Choose DHCP to have an IP address assigned
automatically, or Specify to specify an address.
Press
CHOICE until the desired option is displayed, then
press
ENTER.
Unit[n] SN1234 10 cm3
Set-up Type?
Communications
Unit[n] SN1234 10 cm3
Communications
IP Address?
DHCP