Further settings
93
RS-232
– Print reports
Subsequent selection of cycle reports (see "Process
documentation").
– Language
Any one of the following languages can be set for the RS-232
interface:
German, English (GB), French, Italian, Spanish, Portuguese,
Swedish or Russian.
– Mode
– Terminal
Connection to a terminal or terminal emulator.
Cyrillic characters are not available as ASCII code. When
Russian is selected as the language, the information appears in
English(GB).
– Printer
Connection to log printer.
– Baud rate
Transfer speed of the interface.
– 2400, 9600, 19200, 38400, 57600, 115200.
– Parity
Ensuring data transmission. The parity of the sender and receiver
must match.
– None, Even, Odd.
The following parameters are pre-configured:
Baud rate 9600
Bit 8
Parity None
Stop bits 1