Functional Tests Installation Guide
2 - 32 046-001141-00 A5/A3™ Service Manual
FIGURE 2-45 100% O2 Sensor Calibration
3. The calibration screen shown below is displayed when Next is selected.
FIGURE 2-46 100% O2 Sensor Calibration
4. The calibration screen shown below is displayed when Next is selected. Set up the machine as
per the instructions on the screen. Wait 2 minutes to ensure that the O2 cell voltage has
stabilized at the maximum value for at least 30s. Select Begin.