EasyManuals Logo
Home>Mindray>Medical Equipment>A5

Mindray A5 Service Manual

Mindray A5
436 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #178 background imageLoading...
Page #178 background image
Performance Verification Periodic Maintenance
3 - 38 046-001141-00 A5/A3â„¢ Service Manual
3.11.6 VCV Child Ventilation Mode Test
1. Set the ventilator controls to:
2. Press Set Mode button to begin ventilation.
3. Verify that the pressure waveform, Tidal Volume, Mean or Plateau Pressure, Resp. rate and minute
volume values appear on the screen.
4. Verify the Tidal Volume display is within 18ml of the set value within approximately 1 minute
from the start of ventilation.
5. Verify the delivered volume, as measured by a Vent Tester at the expiratory port, is within 15ml of
the set value within approximately 1 minute from the start of ventilation.
3.11.7 Airway Disconnect Alarm Test
1. While the ventilator is running, disconnect the expiratory limb from the Expiratory Port on the
Breathing System.
2. Verify the following airway pressure disconnect alarm signals activate:
• Paw Too Low message appears on the screen.
• An alarm tone sounds.
Ventilator Controls Ventilator Settings
Ventilation Mode VCV
Vt 120
Rate 15
I:E 1:2
Tpause 10
PEEP Off
Plimit 40

Table of Contents

Other manuals for Mindray A5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A5 and is the answer not in the manual?

Mindray A5 Specifications

General IconGeneral
BrandMindray
ModelA5
CategoryMedical Equipment
LanguageEnglish

Related product manuals