A5/A3â„¢ Service Manual 046-001141-00 2 - 51
Installation Guide Miscellaneous Tests
2.8 Miscellaneous Tests
2.8.1 Test the Line Voltage Alarm
1. Interrupt AC line voltage.
2. Verify that the following alarms activate:
• An alarm tone sounds.
• Battery in use message appears on the screen.
3. Plug the anesthesia system into AC line voltage.
4. Verify that the alarm signals cease.
5. Verify the presence of the battery charging icon in the upper right corner of the screen.
2.8.2 Top Light and Auxiliary Light Test
1. Turn on the Top light located on the bottom side of the top panel.
2. Verify that it lights in both on positions.
2.8.3 Touchpad /USB Mouse Test
For A5 units: Verify that the touchpad is functional.
For A3 units: Verify that the USB mouse can be used to operate the unit.
2.9 Vaporizer Interlock Test
1. Attach two vaporizers to the Vaporizer Mounting Manifold and lock them in place.
2. Rotate either of the vaporizer dial to 3% agent.
3. Verify that the other vaporizer dial cannot be rotated to a setting.
4. Set both vaporizer dials to 0.
5. Rotate the other vaporizer dial to 3%.
6. Verify that the first vaporizer dial cannot be rotated.
7. Rotate both vaporizer dials to T and remove both vaporizers.
8. Verify that the locking spring is intact.
9. Reconnect both vaporizers to the Vaporizer Mounting Manifold.

Table of Contents

Other manuals for Mindray A5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A5 and is the answer not in the manual?

Mindray A5 Specifications

General IconGeneral
TypeAnesthesia Machine
BrandMindray
ModelA5
Power Supply100-240 V AC, 50/60 Hz
Ventilation ModesSIMV, PSV
FlowmetersO2, N2O, Air
Vaporizer CapacityUp to 3 vaporizers
Safety FeaturesO2 failure protection, low pressure alarm
Gas ModuleOptional integrated gas module
Integrated Air CompressorOptional
Patient MonitoringIntegrated monitoring with optional modules

Related product manuals