The Breathing System Theory of Operation
1 - 36 046-001141-00 A5/A3â„¢ Service Manual
1.5 The Breathing System
1.5.1 Brief Introduction
The A5/A3 Breathing System supports three types of operational modes: mechanical ventilation,
manual ventilation, and standby. These modes allow the operator to apply proper ventilation strategy
based on the patient’s needs.
The types of flow paths through the breathing system vary with the operating mode or status.
1.5.2 Automatic Mode, Inspiration
When the Auto/Manual switch is positioned at Auto, the system closes the manual ventilation path.
Drive gas pushes down on the bellows. Gas flows from the bellows, through the CO2 absorber
canister, and through the inspiratory check valve to the patient.
During inspiration, fresh gas flows into the inspiratory limb, upstream of the inspiratory check valve.
In volume mode, tidal volume is compensated for variations in fresh gas flow to ensure that the
volume delivered to the patient meets the set value.
In pressure mode, the inspiratory pressure is regulated both in gas flow and airway pressure to ensure
the airway pressure is held at the set inspiratory pressure during the patient inspiration.
FIGURE 1-31 Automatic Mode, Inspiration Diagram

Table of Contents

Other manuals for Mindray A5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A5 and is the answer not in the manual?

Mindray A5 Specifications

General IconGeneral
BrandMindray
ModelA5
CategoryMedical Equipment
LanguageEnglish

Related product manuals