Disassemble the Breathing System Repair and Disassembly
6 - 62 046-001141-00 A5/A3â„¢ Service Manual
FIGURE 6-119
6.3.15 Remove the Front Upper Cover, Median Plate and Front
Lower Cover Assemblies
1. Remove the Back Upper Cover and Back Lower Cover Assemblies as per section 6.3.16.
2. Remove the two screws on the lower cover.
FIGURE 6-120
3. Loosen the six screws on the upper cover.