A5/A3â„¢ Service Manual 046-001141-00 5 - 75
Repair and Troubleshooting Software Update and Software Configuration Activation
5.6 Software Update and Software Configuration
Activation
NOTE: Software upgrade may by required when replacing the CPU Board (P/N:
801-0631-00026-00), the Ventilator Control Board (P/N: 801-0631-
00027-00), the Power Board (P/N: 801-0631-00025-00) or the Fresh
Flow Sensor Board (P/N: 801-0631-00040-00).
1. Connect the Ethernet port of the PC to the Ethernet port of the A5/A3 using the Ethernet
Crossover Cable.
2. Before running Mindray Patient Monitor System Update Tool, verify that the IP address of the PC
is set to 192.168.23.1., and the Subnet mask is set to 255.255.255.0. To check and set the IP
address on the PC follow these instructions.
1. On the PC click Start, Settings and then Network Connections
2. Right click Local Area Connection and then left click Properties
3. Scroll down to Internet Protocol (TCP/IP), click on it and then click Properties
4. Click the radio button for "Use the following IP address:" Set the IP address and Subnet mask
and then click OK.

Table of Contents

Other manuals for Mindray A5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A5 and is the answer not in the manual?

Mindray A5 Specifications

General IconGeneral
TypeAnesthesia Machine
BrandMindray
ModelA5
Power Supply100-240 V AC, 50/60 Hz
Ventilation ModesSIMV, PSV
FlowmetersO2, N2O, Air
Vaporizer CapacityUp to 3 vaporizers
Safety FeaturesO2 failure protection, low pressure alarm
Gas ModuleOptional integrated gas module
Integrated Air CompressorOptional
Patient MonitoringIntegrated monitoring with optional modules

Related product manuals