Software Update and Software Configuration Activation Repair and Troubleshooting
5 - 76 046-001141-00 A5/A3™ Service Manual
FIGURE 5-73
3. Make sure that the Mindray Patient Monitor System Update Tool has been installed to the PC. If it
has not been installed then follow these steps:
1. Run the SystemUpdateToolForService.exe file (G-110-001396-00).
2. When prompted to setup a language select “English and then select “OK”.
3. When the Welcome dialog window is displayed select “Next.
4. When the Customer Information dialog window is displayed enter the following:
5. User Name: Manufacturing
6. Company Name: Mindray
7. Serial Number: 366-267-2667
8. Select “Next”.
9. When the Administrator password dialog window is displayed enter datascope” as the
password and confirm it, then select “Next”.
10.When the Destination Location dialog window is displayed select “Next” to accept the
default Destination Folders.
11.When the Select Program Folders dialog window is displayed select “Next to accept the
default Program Folder.
12.When the Install Shield Wizard dialog window is displayed select “Finish to complete the
installation.
4. Upgrade Software:
1. From the Desktop, run the Mindray Anesthesia Machine and Ventilator Software Upgrade
Tool 4.3 Icon. When the “Select Product Series dialog is displayed select A Series followed by
OK. Once the System Update Tool starts perform the following to update the software:
2. Select “Select Package from the top tool bar.
3. When the Select Package dialog is displayed select “>>>”

Table of Contents

Other manuals for Mindray A5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A5 and is the answer not in the manual?

Mindray A5 Specifications

General IconGeneral
TypeAnesthesia Machine
BrandMindray
ModelA5
Power Supply100-240 V AC, 50/60 Hz
Ventilation ModesSIMV, PSV
FlowmetersO2, N2O, Air
Vaporizer CapacityUp to 3 vaporizers
Safety FeaturesO2 failure protection, low pressure alarm
Gas ModuleOptional integrated gas module
Integrated Air CompressorOptional
Patient MonitoringIntegrated monitoring with optional modules