A5/A3â„¢ Service Manual 046-001141-00 6 - 3
Repair and Disassembly Disassemble the Assemblies
6.2 Disassemble the Assemblies
6.2.1 Disassemble the Internal Assemblies of the Machine
Upper Half
6.2.1.1 Open the Service Door
Turn the two screws on the service door counter-clockwise one half turn to open the door.
FIGURE 6-1
6.2.1.2 Remove the Gas Supply Inlet Assembly
1. Open the service door.
2. Disconnect the tubes from the N2O, AIR and O2 supply inlet assemblies.
3. Unscrew the two screws from the N2O and AIR supply inlet assemblies.
4. Unplug the related cables from O2 supply inlet assembly and unscrew the two screws to
remove the assembly.
FIGURE 6-2