Mindray A5 - Pressure Support (PS) Ventilation Mode Test

Mindray A5
436 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Verification Installation Guide
2 - 48 046-001141-00 A5/A3™ Service Manual
2.6.10 Pressure Support (PS) Ventilation Mode Test
1. Set the O2 flow to 1 L/min and set the N2O and AIR flow rates to minimum flow.
2. Set the ventilator controls to:
3. Press Set Mode button to begin ventilation.
4. Begin triggering breaths by slightly squeezing the test lung and releasing. Maintain a
continuous breath rate.
5. Verify that a pressure waveform and all ventilation parameters appear on the screen.
6. Verify that the Peak Pressure reading on the display is ±2 the value of
ΔP + PEEP.
7. Stop triggering breaths.
8. Verify that after 15 seconds the ventilator delivers a breath and displays the message Apnea
Ventilation.
9. Verify the system ventilates with a frequency of 4 bpm.
Ventilator Controls Ventilator Settings
Ventilation Mode PS
Min Rate 4
ΔP 20
Trigger 3
PEEP Off
Tslope 0.2
Apnea Ti 5.0