A5/A3â„¢ Service Manual 046-001141-00 6 - 49
Repair and Disassembly Disassemble the Breathing System
6.3.3 Remove the Flow Sensor
1. Turn the locking nuts counterclockwise.
FIGURE 6-93
2. Pull out the inspiration and expiration connectors together with their locking nuts. And then
pull out the flow sensors horizontally.
FIGURE 6-94
3. The following pictures show the appearance of inspiratory and expiratory flow sensor
assemblies.