Performance Verification Periodic Maintenance
3 - 36 046-001141-00 A5/A3â„¢ Service Manual
3.11 Performance Verification
NOTE: Set the Gas Flow Analyzer Correction Mode to ambient temperature
and pressure.
3.11.1 Manual Ventilation Test
1. Ensure that the gas pressure for O2, N2O, and AIR is within specifications.
2. Attach a breathing circuit and test lung to the Y-fitting of the breathing circuit.
NOTE: For testing purposes always use a reusable breathing circuit.
3. Power OFF the anesthesia system.
4. Set the mechanical Auto/Manual switch to MANUAL.
5. Set the APL Valve to approximately 30 cmH2O.
6. Set the AIR flow to approximately 5 L/min using the flow control valve. Use O2 if AIR is not
available.This will require turning the main switch to the ON position.
7. Squeeze the breathing bag once every 10 seconds to inflate and deflate the test lung to
approximately 20 cmH2O of pressure.
8. Verify the inflation and deflation of the test lung.
3.11.2 Manual Mode Ventilation Test
1. Power ON the anesthesia system.
2. Attach a breathing circuit.
NOTE: For testing purposes always use a reusable breathing circuit.
3. Attach a test lung to the Y-fitting of the breathing circuit.
4. Perform the start up tests per the on-screen instructions. Ensure successful completion.
5. Set the mechanical Auto/Manual switch to MANUAL. Press the screen for the screen to change
to manual Mode.
6. Set the APL Valve to approximately 25 cmH2O. Push the O2 Flush button to fill the breathing
bag.
7. Set the AIR flow to 1 L/min using the flow control valve. This will change the screen to manual
Mode
8. Squeeze the breathing bag once every 3 seconds.
9. Verify the inflation and deflation of the test lung.
10. Verify that an airway pressure waveform and all numeric values appear on screen during bag
compressions.
11. Stop squeezing the breathing bag and set the APL Valve to the open position (SP).
3.11.3 APNEA Alarm Test
1. While in the Manual Ventilation Mode, stop ventilating the test lung.
2. Verify that the following APNEA alarm signals activate at approximately 30 seconds from the last
bag compression.
• APNEA appears on the screen.
• An alarm tone sounds.

Table of Contents

Other manuals for Mindray A5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A5 and is the answer not in the manual?

Mindray A5 Specifications

General IconGeneral
TypeAnesthesia Machine
BrandMindray
ModelA5
Power Supply100-240 V AC, 50/60 Hz
Ventilation ModesSIMV, PSV
FlowmetersO2, N2O, Air
Vaporizer CapacityUp to 3 vaporizers
Safety FeaturesO2 failure protection, low pressure alarm
Gas ModuleOptional integrated gas module
Integrated Air CompressorOptional
Patient MonitoringIntegrated monitoring with optional modules